Vapotherm Issues Urgent Nationwide Recall Of All Vapotherm 2000i Respiratory Gas Humidification Devices

Date: 1/26/2006
Reason: Some of the devices have been found to contain the Ralstonia species of bacteria.
Distribution: Nationwide.

Need Legal Help?

Does this recall or safety alert affect you or your family?

The Food and Drug Administration (FDA) released the following information:

Contact:
Kevin Thibodeau
(410) 604-3977

Vapotherm, Inc., Stevensville, Maryland, is initiating a nationwide recall of all Vapotherm 2000i Respiratory Gas Humidification devices.  Some of these devices have been found to contain the Ralstonia species of bacteria.  Ralstonia, as with any gram negative organism, may cause infection, sepsis and in most severe cases be life threatening.

Health care practitioners should seek alternative respiratory gas humidification devices.  Any health care facilities that have the Vapotherm 2000i device must return all devices to Vapotherm, Inc.  Instructions for return are listed on our recall information website at http://www.vtherm.com/recall or by calling Vapotherm, Inc. at 1-866-827-6843.  The "Vapotherm 2000i" label is located on the front of the device on the lower right hand corner.  If there is a question in identification of the product please contact Vapotherm for assistance.

This device is used in both the home and in health care institutions for warming and humidifying breathing gases, such as oxygen, delivered by nasal cannula.

The firm first learned that patients were colonized by the bacteria from a Pennsylvania hospital on August 17, 2005, and subsequently issued a voluntary recall of the Vapotherm 2000i on October 13, 2005.  FDA has since been apprised of this action.

At this time, the following information is known:

There are numerous reports of Ralstonia colonization, including three reports of infection.
One hospital, reported a death, but this has not been confirmed by Vapotherm
26 hospitals in 16 states have reported positive cultures of Ralstonia species from the Vapotherm 2000i device. 
Vapotherm's investigation is currently ongoing to identify the source of the Ralstonia contamination.  In the meantime, Vapotherm's plans include recalling and performing a disinfection process on the units. 

The Vapotherm 2000i devices were distributed through specialty distributors in the U.S. and internationally.  Consumers with questions regarding this recall may contact the company at 1-866-827-6843, or Kevin Thibodeau at 410-604-3977.

Any adverse reactions experienced with the use of this product, and/or quality problems should also be reported to the FDA's MedWatch Program by phone at 1-800-FDA-1088, by Fax at 1-800-FDA-0178, by mail at MedWatch, HF-2, FDA, 5600 Fishers Lane, Rockville, MD 20852-9787, or on the MedWatch website at www.fda.gov/medwatch.

Type of Recall:
Class I - This is a health hazard situation where there is a reasonable probability that the use of the product will cause serious, adverse health consequences or death.


Other Drugs / Medicines Recalls and Alerts of Interest

Become a LawCash Member - FREE!
'Find Money' E-Book
Weekly Email Alerts




privacy policy
YouNewz Beta
IT'S FREE

Report

Report Newz and easily upload your own newzworthy photos from your cell phone or computer to the web.

Share

Quickly share your photos with family, friends, co-workers, or the world with your own Newzpaper.

Read

Instantly find Newz and photos from other YouNewzers and read other YouNewzers Newzpapers.
 
Class Action Lawsuit Center || Product Recall Center || Consumer Complaint Center || About LawCash Link Exchange
Privacy Policy || Legal Policies || Terms & Conditions || Website Advertising Policy || Site Map || Top Lawsuits
LawCash® is a service of nola3, llc
© 2000 - 2008 Copyright. All rights reserved nola3, llc.

[ Home ]
LawCash
login
Justice is a click away.