Disetronic Medical Systems, Inc. Announces a Voluntary Nationwide Recall Of ALL ACCU-CHEK Ultraflex Infusion Sets

Date: 4/5/2006
Reason: Tubing could fully or partially separate at the luer lock-tubing connection
Distribution: Nationwide.

Need Legal Help?

Does this recall or safety alert affect you or your family?

The Food and Drug Administration (FDA) released the following information:

Disetronic Medical Systems, Inc, (Disetronic) of Fishers, Ind. announced a voluntary nationwide recall of ALL ACCU-CHEK Ultraflex Infusion Sets, because of a potential that tubing could fully or partially separate at the luer lock-tubing connection. In the event that a full or partial separation occurs, it is possible that insulin could leak from the infusion set tubing causing an interruption of insulin delivery, which can cause hyperglycemia.

This recall applies to ALL ACCU-CHEK Ultraflex infusion sets. Patients using any standard luer-lock insulin pump may also be using these ACCU-CHEK Ultraflex infusion sets. Disetronic is advising customers to check their infusion sets at the luer lock-tubing connection at least every 3 hours and before bedtime.

Under this recall, customers have the option of continuing to use their ACCU-CHEK Ultraflex infusion set, and receiving replacement ACCU-CHEK Ultraflex infusion sets for any products exhibiting full or partial separation of the luer lock-tubing connection. Customers deciding to continue using the ACCU-CHEK Ultraflex or a replacement ACCU-CHEK Ultraflex infusion set, must check their infusion set at the luer lock-tubing connection during use at least every 3 hours and before bedtime.   The replacement ACCU-CHEK Ultraflex infusion sets may still experience this full or partial separation of the luer lock-tubing connection.

Customers also have the option of replacing their ACCU-CHEK Ultraflex infusion sets with ACCU-CHEK Tender, or ACCU-CHEK Rapid-D infusion sets.

Customers that have fully or partially separated tubing sets, or wish to discontinue use of their ACCU-CHEK Ultraflex Infusion Set may call Disetronic Medical Systems Pump Support at 1-800-688-4578 for replacement. If you do experience product problems, you should contact Disetronic, for a pre-paid mailer in order to send the affected infusion set back to Disetronic for analysis.

Disetronic is conducting this recall of ACCU-CHEK Ultraflex infusion sets now, because of a recent increase in complaints regarding fully or partially separated luer lock-tubing connections. Disetronic has an ongoing investigation into these issues and will update our customers, their healthcare providers and our distributors as new information becomes available. If you are a physician or a patient who has experienced a problem with any ACCU-CHEK Infusion Sets, please notify Disetronic at 1-800-688-4578.

Infusion set customers who purchased the ACCU-CHEK Ultraflex Infusion sets should always follow labeling instructions for proper use. In addition, healthcare officials suggest that diabetes patients who are on insulin pump therapy should check their blood glucose levels 1-3 hours after changing the infusion set systems, and continue to check throughout the day as instructed by their healthcare provider.

In the course of normal daily use, infusion sets can be subjected to a variety of stresses, including bending at the luer lock-tubing connection, which can lead to full or partial separation at the luer lock-tubing connection. Therefore, the potential for separation of the luer lock-tubing connection still exists with the replacement ACCU-CHEK Ultraflex infusion set. Information and illustrations that depict the luer lock-tubing connection can be found on Disetronic's website at www.disetronic-usa.com.

ACCU-CHEK Ultraflex Infusion Sets are available by prescription only for diabetes patients who use insulin pump therapy. The firm is notifying by direct mail affected customers, healthcare providers and its distributors of this action.

Any adverse reactions experienced with the use of this product, and/or quality problems should also be reported to the FDA's MedWatch Program by phone at 1-800-FDA-1088, by Fax at 1-800-FDA-0178, by mail at MedWatch, HF-2, FDA, 5600 Fishers Lane, Rockville, MD 20852-9787, or on the MedWatch website at www.fda.gov/medwatch.

"After learning about this problem and our records of complaints, we decided to voluntarily notify patients, healthcare providers and distributors regarding this product. We care about patient safety, their well being and peace of mind and will be pleased to work with anyone who desires a replacement of their ACCU-CHEK Ultraflex infusion set." said Sarah Hanssen, Vice President of Disetronic Medical Systems. "We are working with the FDA to ensure the proper notification of patients and healthcare providers."

 


Other Drugs / Medicines Recalls and Alerts of Interest

Date: 6/12/2008
Reason: it contained a potentially harmful undeclared ingredient, hydroxyhomosildenafil

Distribution: Nationwide
 
Date: 6/12/2008
Reason: Limited lots of Carmeda-coated products were manufactured with heparin found to have been contaminated with oversulfated chondroitin sulfate (OSCS).

Distribution: Nationwide
 
Date: 6/16/2008
Reason: Oversized tablets may contain as much as two times the labeled level of active morphine sulfate

Distribution: Nationwide
 
Date: 6/12/2008

Reason: Oversized tablets may contain as much as two times the labeled level of active morphine sulfate

Distribution: Nationwide

 
Date: 6/12/2008
Reason: Limited lots of Atrium heparin-coated Hydraglide Thoracic Catheters were manufactured with heparin found to have been contaminated with oversulfated chondroitin sulfate (OSCS)

Distribution: Nationwide
 
Date: 6/12/2008

Reason: Taking AZASAN® instead of SOLODYN® presents a health hazard and safety risk to patients

Distribution: Nationwide

 

Become a LawCash Member - FREE!
'Find Money' E-Book
Weekly Email Alerts




privacy policy
YouNewz Beta
IT'S FREE

Report

Report Newz and easily upload your own newzworthy photos from your cell phone or computer to the web.

Share

Quickly share your photos with family, friends, co-workers, or the world with your own Newzpaper.

Read

Instantly find Newz and photos from other YouNewzers and read other YouNewzers Newzpapers.
 
Class Action Lawsuit Center || Product Recall Center || Consumer Complaint Center || About LawCash Link Exchange
Privacy Policy || Legal Policies || Terms & Conditions || Website Advertising Policy || Site Map || Top Lawsuits
LawCash® is a service of nola3, llc
© 2000 - 2008 Copyright. All rights reserved nola3, llc.

[ Home ]
LawCash
login
Justice is a click away.